Technical Data
- Biometra TRobot II
PCR module
| Biometra TRobot II 96 G | Biometra TRobot II 96 SG | Biometra TRobot II 384 G |
Sample block | Aluminum, special alloy | Silver, gold coating | Aluminum, special alloy |
Block capacity | 96 x 0.2 mL tubes, 96 well microplate or 8 well strips | 96 x 0.2 mL tubes, 96 well microplate or 8 well strips | 384 well microplate |
Proposed sample volume | 5–50 µL | 5–100 µL | 5–25 µL |
Maximum heatinga | 4.0 °C/sc | 5.0 °C/s | 2.4 °C/s |
Average heatinga | 3.7 °C/sc | 4.9 °C/s | 2.2 °C/s |
Maximum coolinga | 3.1 °C/sc | 4.8 °C/s | 1.8 °C/s |
Average coolinga | 2.8 °C/sc | 4.5 °C/s | 1.6 °C/s |
Temperature uniformityb | 95 °C ± 0.60 °C | 95 °C ± 0.50 °C | 95 °C ± 0.60 °C |
Gradient | Linear Gradient Tool | Linear Gradient Tool | Linear Gradient Tool |
Max./Min. gradient | 24 °C/0.1 °C | 30 °C/0.1 °C | 18 °C/0.1 °C |
Adjustable gradient range | 12 rows 20 °C to 99 °C | 12 rows 10 °C to 99 °C | 24 rows 20 °C to 99 °C |
Block exchange | No | ||
Tempering method | Peltier elements | ||
Standby temperature | Yes, down to 4 °C | ||
Temperature control mode | Block control | ||
Adjustable temperature range | 3 °C to 99 °C | ||
Temperature control accuracy | ± 0.1 °C | ||
Lid type | Motorized lid with heating module, which is connected spring-loaded to the lid | ||
Lid temperature | 30 °C to 110 °C | ||
Contact pressure | Constant contact pressure independent from the used consumables adjustable via software in the range of 4 kg to 12 kg. Typical standard settings: 8 kg to 10 kg | ||
Compatible sealing systems | Sealing foils, mats and lids | ||
Plate lifter | Motor-driven, direct compatible with full-skirted PCR plates. With optional Biometra TRobot II adapter frame also compatible with semi-skirted PCR plates with the following plate dimensions: | ||
Consumable compatibility | Low-profile and high-profile plastics with or without skirt, as well for semi-skirt | ||
Application software | Biometra TSuite computer software | ||
Integration software | Biometra Library (dll file) for integration into an automation system´s scheduler software | ||
Operating system | Microsoft Windows minimum Windows 7, preferably Windows 10 | ||
Language | English, German | ||
Data transfer | Via Ethernet, USB2.0 or serial RS232 interface (encrypted transmission of data packets for run log files and service files) | ||
Autorestart function | Yes | ||
Quick start function | Via dashboard in Biometra TSuite software | ||
Time increment | 1 to 240 s/cycle | ||
Temperature increment/decrement | ±0.1 to 20 °C/cycle | ||
Memory capacity | Controller: 394 programs of 6 steps in up to 90 user folders. On computer or network: unlimited, depending on storage space | ||
Further features |
| ||
Noise emission | Max. 45 dB(A) (measured for standard applications) | ||
Operating conditions | 15 °C to 30 °C, max. 70 % humidity, max. 2000 m above sea level, overvoltage category II, pollution degree 2, IP20 | ||
Dimensions | 208 mm x 260 mm x 209 mm (closed) 208 mm x 332 mm x 264 mm (with open lid) 208 mm x 315 mm x 291 mm (with open lid in middle opening position) | ||
Required clearance zone | 20 cm to the rear of the device. When operating several units side by side, an additional 10 cm between the units. | ||
Weight netto | 9.5 kg | ||
Interfaces to the controller | For DSub 24W7 signal cable and for DSub 12 supply cable. | ||
Warranty | 2 years |
a measured at cavity wall of the block
b after 15 s
c preliminary data
Controller
Power supply | 100 V, 115 V or 230 V ±10 %, 50 – 60 Hz | |
Fuse | 2x 8AT 250 V (specific fuse type needed). Contact the Analytik Jena Service. | |
Power consumption | Max. 550 W | |
Interfaces to the computer | Serial RS232, USB2.0 via RS232-to-USB adapter and Ethernet. A 3 m long RS232 cable is included in delivery. The network cable must be at least performance class Cat 5e and the cable configuration has to be STP. Cable lengths ≤ 30 m are permitted. | |
Interfaces to the PCR module | DSub 24W7 signal cable, DSub 12 power cable. Cable lengths: 3 m. The cables are permanently connected to the controller. | |
Dimensions (W × D × H) | 240 mm x 397 (444 incl. cable ports) mm x 139 mm | |
Weight | 8.4 kg | |
Required clearance zone | 10 cm to the rear of the device. |